• HC-CK101 Crack Width Observer
HC-CK101 Crack Width Observer
+
  • Product Describe
  • Product Configuration

    Detector Host Camera Microscope Probe Signal Connection Cable Calibration Scale Plate Charger Instrument Case

    Usage Instructions

    Connect the measuring probe and the instrument host with the dedicated transmission cable. After turning on the host, place the tip of the measuring probe close to the crack to be measured. The host display will show the magnified crack; fine-tune the position of the measuring probe so that the crack is perpendicular to the screen scale line, then interpret the actual width of the crack based on the crack image.

    Instrument Calibration

    The 2mm scale line of the standard scale can be seen on the screen. When adjusting the measuring probe position for calibration, first align the measuring probe with the two reference lines on the upper and lower edges of the calibration scale plate. If the magnified image of the standard 2mm scale line coincides with the 2mm scale line on the screen and the error does not exceed 0.02mm, it indicates that the instrument's magnification is within the normal range and can be used normally. If the error exceeds 0.02mm, please contact us.

    Technical Parameters

    Measurement Range: 0.01mm ~ 2.00mm
    Reading Accuracy: 0.005mm
    Magnification: 40x
    Minimum Scale Division: 0.02mm
    Power Supply: Built-in rechargeable lithium battery, can work for 7 hours
    Screen Size: 109x65mm
    Host Size: 135x101x30mm
    Probe Size: 55x50x54mm
    Cable Length: 1.5m
    Total Weight: 650g

    Product Configuration

    1. HC-CK101 Crack Width Detector Host (Screen Scale Line)
    2. Microscope Camera Measuring Probe
    3. Battery Charger
    4. Signal Transmission Cable
    5. Scale Calibration Plate

    Precautions

    1. Please charge before first use. The charger indicator light will turn green when charging is complete.

    2. Before use, check the scale line on the calibration plate. The error between the magnified 2mm image and the 2mm on the screen should be less than 0.02mm.

    3. When measuring, try to keep the crack image perpendicular to the screen scale line.

    4. Keep the measuring lens clean. Only use an air blower or soft brush for cleaning.

    5. The instrument undergoes strict calibration before leaving the factory and generally does not require self-adjustment of the camera probe. If the error exceeds 0.02mm, please contact the manufacturer for calibration.

    6. The number of indicator lights lit on the upper right of the instrument represents the remaining battery power, each segment equals 25%.

    7. The instrument will automatically shut down after being idle for more than ten minutes to save power.

    Troubleshooting

    Fault 1: After powering on, the host screen shows no display → Check if the battery is installed properly and if the battery is sufficiently charged. → Check if the connection between the host and the measuring probe is stable and inspect the cable for any damage.
    Fault 2: When observing cracks, the edges of the crack image are blurry → Clean the surface of the crack to be measured to make it relatively flat.

Company Profile

Hebei Yinfeng Experimental Instrument Co., Ltd. is a high-tech enterprise dedicated to the research and development, production, and sales of experimental instruments. The company is headquartered in Hebei Province, relying on the strong industrial foundation and technological innovation resources in the Beijing Tianjin Hebei region. It is committed to providing high-precision and high reliability testing equipment and solutions for material testing, engineering quality control, scientific research experiments and other fields.

Plant
Factory environment
Workshop
Workshop environment

Customized Delivery Process

We provide customers with full-process services ranging from pre-sale consultation, customized solution design, equipment installation and commissioning to after-sale technical support.

Online Communication

Provide Custom Drawings

Merchant Quotation

Sign A Contract

Processing And Production

Packaging And Distribution

Confirm Receipt Of Goods

Successful Transaction

  • HC-CK101 Crack Width Observer
HC-CK101 Crack Width Observer
+

HC-CK101 Crack Width Observer

Connect the measurement probe and the main instrument using a dedicated transmission line. After turning on the main instrument, place the tip of the measurement probe close to the crack being measured. The main instrument's display will show the magnified crack; fine-tune the position of the measurement probe so that the crack is perpendicular to the screen's scale line, then interpret the actual width of the crack based on the crack image.

If you need customized products,  Contact US !

Share to

Tag list

Request a Quote

We will contact you within one working day. Please pay attention to your email.

Related Products

Content update in progress

Request Quote

We will contact you within one working day. Please pay attention to your email.

SUBMIT